FOCAL2 (acronym of Facility for Optical Calibration At Liège) is a cylindrical stainless steel vessel of 17m3 with an horizontal axis and a diameter of 2m. Its length is 4.52 m.

Characteristics

Access flanges in the chamber

IMG 0450s

 

 

 

 

 

 

 

 

For light beam access, thermal fluid transfer and electrical supply, 22 flanges (mainly of 200 mm diameter) are distributed over the whole chamber surface.

Focal 2 optical bench

IMG 4079s

 

 

 

 

 

 

 

 

An optical bench is installed in the vacuum chamber. It is supported by 4 feet passing through the wall of the vessel and is insulated from the chamber mechanical vibrations by bellows, which also ensure vacuum tightness.

These feet stand on a massive seismic block

The seismic block is larger than the vacuum chamber (length: 8.67m) allowing to put the OGSE outside chamber.

The optical bench sizes are 4.7m by 1.3m. A threaded holes network (M10 separated of 300 mm) is foreseen on the upper surface. 

The optical bench can be moved out of the chamber by means of a railway system.

  IMG 0451s   IMG 0066s

Pumping system

The pumping cycle is divided into 3 steps:

  • dry primary pumping
  • dry turbomolecular pumping
  • dry interseal pumping

This pumping system is designed to reach 5 10-3 Pa after 3 hours of pumping (this value is given for an empty chamber).

Crane

The 5T clean room crane can be used to install the test set-up and the specimen on the Focal2 optical bench.

Thermal system

Thermal environments inside the chamber can be simulated by means of dedicated shrouds, fed with nitrogen and/or helium.

  • NITROGEN thermal system :

IMG 4102s

 

 

 

 

 

 

 

 

5 independent thermal channels allow to feed thermal panels. The temperatures can be set in the range from 123 [K] to 353 [K] for the regulation system or 85 [K] for the liquid nitrogen system (each line is equipped with a valve which selects GN2 or LN2 mode).

It is possible to reach higher temperature than 353 [K] with additional heaters glued on the specimen or with Infrared lamps installed inside the thermal tent.

The thermal regulation system is designed to avoid oscillations and temperature drift.

It warrants performances about +/-0.1 [K] in long term stability and allows complex thermal cycling tests.

  • HELIUM thermal system :

The Helium thermal system is composed of lines fed by one of the Helium liquefier/refrigerator (KOCH, model 1630  or LINDE TCF 20) available at CSL. One Focal2 line can be fed with Helium.

For the cold gaseous Helium line, the temperature can be controlled from 300 [K] to 20 [K].

It should be noted that the system architecture allows to feed the lines with warm Helium up to 353 [K].

  • Nitrogen and helium connection :

The fluid connection for piping is SS-8-VCR Cajon type.

Data Handling

The data acquisition of Focal 2 is organized  Keithley 2701 data loggers with industrial PC’s.

The current sensor capacity is 90 thermocouples (2 wires) or 45 PT100 (4 wires) but the number can be adapted.

The here below type of sensors are usable and readable:

Thermocouple : T and K type Resistance 4 wires
Diode 4 wires Cryogenic diode
PT100 2 wires Cernox ([Ω] or [K])
PT100 4 wires Thermistor
PT500 2 wires Voltage 0-1 [V] et 0-10 [V]
PT1000 2 wires Current< 100 [mA]
PT2000 4 wires  

 

Some display screens are available in the command room and in the customer areas. Temperatures, gradients, alarms and warnings can be recorded and plotted.

CSL provides various communication means between the data handling system and the customer:

  • The customer can be connected to the Facilities Ethernet network or have a connection to his home base through the public Internet, or through a dedicated ISDN link.
  • "Teletesting",i.e the real time monitoring and control of an instrument by a customer remote data logger, can also be implemented.

Clean room and cleanliness aspects

IMG 4315s

The Focal 2 vacuum chamber is located in a class 10 000 (ISO 7) clean room (following US standard:FED-STD-209E).To satisfy specific requirements, a class 100 ( ISO 5) of 100 m3 (28m2) has been built around Focal 5. Movable parts allow to use the clean room 5T crane.

 Plan f2

class 100 F2